| Symptom |
Analyze |
Remedy |
| Mineral Build Up Inside Chamber |
Type Of Water That Is Being Used |
Clean Inside Of Chamber Thoroughly - Refill Using Distilled Water |
| Sterilizer Takes Over 15-20 Minutes (From a Warm Start) To Reach Set Sterilization Temperature |
Chamber Load Bellows
Line Voltage
Solenoid Valves |
Place Fewer Packages Inside Chamber Replace If Necessary (Part # PCB001)
If Line Voltage Is Too Low Plug Into Outlet With Proper Voltage - Contact Qualified Electrician.
Refer To Panel 5 (Water Reservoir Leak Identification To Identify LeaksClean Inside Of Chamber Thoroughly |
| Displayed Temperature Stays Considerably Below Required Temperature For Sterilization Cycle |
Ribbon Cable
Bellows
Dump Valve. Refer To Panel 5 (Water Reservoir Leak Indentification) |
If Necessary Replace Ribbon Cable (Part # PCC108)
If Necessary Replace Bellows (Part # PCB001)
If Necessary Clean Dump Valve And/Or Replace Valve Kit (Part # PCK106). See Panel 8
|
| Wet Packs After Drying Cycle Is Complete Or Wet Packs Between Layers |
Chamber Load
Fill-Line Filter Condition & Position On Chamber Floor |
Break Sterization Load Into Fewer Or Smaller Packs/Wraps For Proper Steam Sterilization.Crack Door Open - 1"
During Drying Cyle.
Replace Fill-Line Filter And/Or Adjust Filter To Center Bottom Of Chamber (Part # PCF009) |
| Water In Bottom Of Sterlizer When Unit Is Not In Fill Cycle |
Fill Valve |
Clean Fill Valve And/Or Replace Valve Kit (Part # PCK102). See Panel 8 |
| Water Dripping From Door When Door Is Opened After Cycle |
Spacer (Door) |
Check Spacer For Shrinkage Or Damage. Replace If Necessary (Part # PCS 067) |
| When Unit Vents, Water Comes Out Of Fill Port On Top Of Case |
Water Level Inside Reservoir |
Drain Water To The Bottom Holes Of Fill Cup |
| Sterilizer Fails Several Self-Tests, But Still Operates |
Cable Connections |
Make Sure Cable Connectors At Front Panel And Power Board Are Good |
| Sterilizer Does Not Turn On |
Cable And On/Off Switch (Power LED On Power Supply Board Should Be Illuminated) |
Check MicroProcessor Board On/Off Switch And Cable With Ohm Meter To Verify Continuity*
If Switch & Cable Pass Test, Replace Power Supply Board |
| Sterilizer Does Not Switch Modes |
Gap Between Panel Membrane And Switches Mounted On Micro-Processor Board |
Check Between MicroProcessor Board And Panel Membrane For Washers Or Materials Causing Incorrect
Separation. Replace MicroProcessor Board If Not Correctable. |